41
Starting Measurement
Trigger measurement operation (in the STOP state or when the trigger source is
set to EXTERNAL
;withcontactchecksettoON)
When the contact check function is set to OFF, no contact check will be performed after trigger
input.
Example1:Measurementcountof1
TRIG ON OFF
EOM ONOFFON
Wait for trigger Trigger standby
Contact
check
Delay Measurement
Example2:Measurementcountofn
TRIG ON OFF
EOM ONON OFF
Measurement time
Measurement count = n
Contact
check
Delay
1
st
measurement
2
nd
measurement
n
th
measurement
Trigger
standby
Measurement time (reference value)
Integration time setting Measurement time [ms]
0.02PLC 0.4 × n
0.2PLC (50 Hz) 4 × n, (60 Hz) 3.2 × n
1PLC (FAST) (50 Hz) 20 × n, (60 Hz) 16.7 × n
10PLC (MEDIUM) (50 Hz) 200 × n, (60 Hz) 167 × n
100PLC (SLOW) (50 Hz) 3900 × n, (60 Hz) 3400 × n
ms Integration time × n
Measured values up to the (n-1)
th
measurement are not used in comparator or BIN judgment.
Only the n
th
measured value is used in judgment output.
Wait for trigger
3
Measurement