137
Conguring External Input and Output
TRIGsignaloperationwhiletheinputlterissettoON
(Chatter)
Response
time
During measurement
Hold the input signal until the response time has passed.
EOMsignaloutputtype
You can choose to either hold the EOM signal’s output until the next trigger is received or output
the set pulse.
(Measurement screen) MENU >
EXTI/O
1
2
Selecttheoutputtype.
HOLD Holds the EOM signal after
measurement is complete
(default setting).
PULSE Outputs a pulse with the set
width after measurement is
complete.
3
(After selecting PULSE)
Set the pulse width and touch ENTER.
˄ Increases the value by 1.
˅ Decreases the value by 1.
CLR Deletes all characters.
CANCEL Cancels the setting and returns to
the previous screen.
Valid setting range: 1 ms to 100 ms
(Default setting: 5 ms)
Measurement
processing
External Control (EXT I/O)
11
w ww . . co m
information@itm.com1.800.561.8187