EasyManua.ls Logo

Hioki DM7275 Series - X84; to Perform Judgment after Measured Values Stabilize; X84; to Output Judgment Results to an External Device or to Print Judgment Results

Hioki DM7275 Series
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56
Comparator Measurement (Using a Single Judgment Standard)
Toperformjudgmentaftermeasuredvaluesstabilize
Since measured values may exhibit instability immediately after the instrument is connected to the
measurement target when performing measurement manually, readings may temporarily exceed
the judgment range.
When judgment delay is enabled, the judgment will be output after the same judgment is obtained
the set number of times.
(Measurement screen) MENU > MATH > COMP
1
2
Enablethejudgmentdelay.
(Default setting: OFF)
3
Selectthedesirednumberofjudgment
delays.
(Default setting: 2)
Tooutputjudgmentresultstoanexternaldeviceortoprintjudgment
results
Set the comparator function to ON, congure external output (p. 111) or printing (p. 143), and
prepare the associated equipment.