176
Error Displays
14.3 ErrorDisplays
The following messages are displayed on the screen when the instrument malfunctions or
encounters an abnormal measurement state.
• If you feel that the instrument may be malfunctioning, contact your authorized Hioki distributor or
reseller after reviewing the information provided in “Q&A (Frequently Asked Questions)” (p. 168).
• When an error is displayed on the LCD screen and service is required, please contact your
authorized Hioki distributor or reseller.
Display Description Solution
+OvrRng/−OvrRng Over-range Select the appropriate range.
(p. 35)
NoCntct Contact error Check the connections with
the object to be measured.
(p. 31)
Or, adjust the threshold for the
contact check.(p. 71)
Err.TC Temperature compensation
error
Connect a temperature sensor.
(p. 27)
ERR:001
Lowerlimitishigherthan
Upper limit.
Cannot set because the lower
limit value is greater than the
upper limit value.
Set an upper limit value that
is greater than the lower limit
value.
(p. 51)
ERR:004 Unable to change the
setting during auto-hold.
When auto-hold function
is enabled, settings for
measurement speed and
continuous mode cannot be
changed.
Turn OFF the auto-hold
function. (p. 70)
ERR:005
UnabletosetNULL
due to an abnormal
measurement value.
When OverRng, NoCntact, or
------ is displayed, NULL value
cannot be obtained.
Return from the abnormal
measurement state. (p. 46)
ERR:030 Command error. Remote command syntax
error. (String is incorrect or
incorrect character code is
used.)
Check if the commands are
correct. (See the application
disc provided.)
ERR:031
Executionerror.Invalid
parameter.
Remote command execution
error. The parameter value is
out of range.
Check if the parameters are
correct.
ERR:032
Executionerror.
Remote command execution
error.
Check the execution error
conditions for each command.
ERR:050
Thepaneldoesnotexist.
Panels that have not been
saved cannot be read.
Select a proper panel. (p. 61)
ERR:051
Thepaneldoesnotexist.
Unable to rename.
Name of panels that have
not been saved cannot be
changed.
Select a panel that has been
saved. (p. 61)
ERR:060
CannotuseUSBmemory.
SetI/FfunctiontoUSB-
MEMORY.
When I/F is set to USB COM,
USB ash drive cannot be
used.
Set the I/F to USB MEMORY.
(p. 115)
ERR:061
Thedriveisnotready.
(NoUSBmemory
inserted)
USB ash drive is not inserted.
Insert a USB ash drive.
(p. 115)
ERR:062 This format is not
supported.
Format of the USB ash drive
is not correct.
Format the USB ash drive to
FAT32.
w ww . . co m