47
Measurement Error Displays (Displays Other Than Measured Values)
Measurement error detection order
Measurement errors are detected in the order shown in the diagram below. The rst error that is
detected will be displayed on the screen and output as a signal from the EXT I/O.
Order Measurement error
detection
Screendisplay EXTI/Oconnector
1
Temperature
compensation error
Yes
Err.TC ERR signal output
No
2
Display:Higherthan
upper limit
Yes
+OvrRng
HI signal output
(When comparator is ON)
No
3
Display:Lowerthan
lower limit
Yes
−OvrRng
LO signal output
(When comparator is ON)
No
4 Contact error
Yes
NoCntct ERR signal output
No
5 No measurement data
Yes
−−−−−−− No output
3
Measurement
w ww . . co m
information@itm.com1.800.561.8187