180
Disposing of the Instrument
14.4 Disposing of the Instrument
The instrument uses a lithium battery as a backup for its clock.
When disposing the instrument, remove the lithium battery and dispose the battery and instrument
in accordance with local regulations.
WARNING
Toavoidelectricshock,turnOFFthemainpowerswitchanddisconnectthe
powercordandmeasurementcablesbeforeremovingthelithiumbattery.
RemovingtheLithiumBattery
Required tools:
• One Phillips screwdriver (No.1)
• One pair of tweezers (to remove the lithium battery)
1
Verifythatthepowertotheinstrumentis
OFFandunplugthepowercord,andany
other cords or cables.
2
Removethesixscrewsfromthesides.
Lithiumbattery
3
Remove the cover.
4
Insertthetipofthetweezersbetweenthe
batteryandthebatteryholderasshown
inthepictureandliftuponthebatteryto
remove it.
CAUTION
Exercise care not to short the positive
and negative terminals. Doing so may
cause sparks.
CALIFORNIA,USAONLY
This product contains a CR Coin Lithium Battery which contains Perchlorate Material -
special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
w ww . . co m