Error Displays
Display Description Solution
ERR:063 Error while reading the
USBmemory.
An error occurred while
reading the USB ash drive.
The le may be damaged.
Recover the le or use a
different USB ash drive.
ERR:064 Error while reading the
congurationle.
An error occurred while
reading a setting le in the
USB ash drive.
The le may be damaged.
Recover the le or use a
different USB ash drive.
ERR:065 File not found. A valid le was not found in the
USB ash drive.
Specify a proper le.
ERR:070 No space available. There is no free space in the
USB ash drive.
Delete unnecessary les to
secure free space.
ERR:071 Error occurred saving
thele.
An error occurred while saving
data in the USB ash drive.
The le may be damaged.
Recover the le or use a
different USB ash drive.
ERR:076 Error occurred deleting
thele.
An error occurred while
deleting data in the USB ash
drive.
The le may be damaged.
Recover the le or use a
different USB ash drive.
ERR:077
Unabletorenamethele
becauseanotherlewith
thesamenamealready
exists.
The le name cannot be
changed as there are les with
the same le name.
Specify a different le name.
ERR:078 Error occurred renaming
thele.
An error occurred while
changing a le name in the
USB ash drive.
The le may be damaged.
Recover the le or use a
different USB ash drive.
ERR:079 Error while reading the
USBmemory.
An error occurred while
reading the USB ash drive.
The le may be damaged.
Recover the le or use a
different USB ash drive.
ERR:080 Unable to enter the
adjustmentmode.
The mode cannot be changed
to adjustment mode.
The Adjustment screen is not
available for use by end-users.
ERR:090
ROMchecksumerror.
Check sum of the program
ROM does not match.
Malfunction of the instrument.
Request service.
ERR:091 RAM error. The RAM failed. Malfunction of the instrument.
Request service.
ERR:092
Memoryaccesserror.
Turn off the power and
restart after a while.
Communication with memory
failed.
Turn OFF the power and turn it
ON again after some time.
ERR:093
Memorytesterror.
Memory failure. Malfunction of the instrument.
Request service.
ERR:094
Adjustmentdataerror.
Adjustment data is not correct. Malfunction of the instrument.
Request service.
ERR:095
Backupdataerror.
Backup data is not correct. Settings have been reset.
Recongure measurement
conditions and other settings.
ERR:096 Failed to detect line
frequency.Selectline
frequency.
The power frequency has not
been detected.
Check the voltage and
frequency of the power supply.
(p. 90)
ERR:098
"Theclockisnotset.
Reset?(15-01-01
00:00:00)"
Clock has not been set. Replace the backup battery
and set a clock.
ERR:099 Failed to detect line
frequency;willbesetto
50Hz.
The power frequency has not
been detected. The frequency
will be set to 50 Hz.
Check the voltage and
frequency of the power supply.
(p. 90)