Operating Precautions
Before connecting a measurement cable
DANGER
Measurementcablesshouldonlybeconnectedtothesecondarysideofa
breaker.Anyshort-circuitcurrentatthesecondarysidewillbecut-offbythe
breaker.Connectionsshouldneverbemadetotheprimarysideofabreaker,
becauseunrestrictedcurrentowcoulddamagetheinstrumentandfacilitiesifa
short circuit occurs.
WARNING
To avoid electric shock and short-circuit accidents, use only the specied measurement
cables to connect the instrument input terminals (HIGH and LOW terminals) to the
circuit over 70 V DC to be tested.
Before connecting a temperature sensor
IMPORTANT
Connect the temperature sensor by inserting the plug all the way in the TEMP.SENSOR
connector. Insufcient connection may cause larger errors in measured values.
Before turning the power ON
WARNING
Beforeturningtheinstrumenton,makesurethesupplyvoltagematchesthat
indicatedonitspowerconnector.Connectiontoanimpropersupplyvoltagemay
damagetheinstrumentandpresentanelectricalhazard.
CAUTION
Avoid using an uninterruptible power supply (UPS) or DC/AC inverter with rectangular
wave or pseudo-sine-wave output to power the instrument. Doing so may damage the
instrument.