Operating Precautions
BeforemakingaconnectiontotheUSBconnector
CAUTION
• To avoid equipment failure, do not disconnect the USB cable while communications
are in progress.
• Use a common ground for both the instrument and the computer. Use of different
ground circuits will result in a potential difference between the instrument's ground
and the computer's ground. If the USB cable is connected while such a potential
difference exists, it may result in equipment malfunction or failure.
BeforeconnectingtotheRS-232CorGP-IBconnectors
CAUTION
• Use a common ground for the instrument, computer and programmable controller.
Use of different ground circuits will result in a potential difference between the
instrument's ground and grounds of the computer and programmable controller. If the
communications cable is connected while such a potential difference exists, it may
result in equipment malfunction or failure.
• After connecting the communications cable, tighten the screws on the connector
securely. Failure to secure the connector could result in equipment malfunction or
damage.
BeforeconnectingaUSBashdrive
CAUTION
• Inserting a USB ash drive upside down, backwards or in the wrong direction may
damage the USB ash drive or the instrument.
• Some USB ash drives are susceptible to static electricity. Exercise care when using
such products because static electricity could damage the USB ash drive or cause
malfunction of the instrument.
With some USB ash drives, the instrument may not start up if power is turned on while the USB
ash drive is inserted. In such a case, turn power on rst, and then insert the USB ash drive.
It is recommended to try out operation with a USB ash drive before starting to use it for actual
measurements.
Beforeswitchingthecurrentsink(NPN)/currentsource(PNP)
CAUTION
• Never switch between NPN and PNP while the instrument's power is on.
• Congure the NPN/PNP setting based on the externally connected device.
10
9
8
7
6
5
4
3
2
1
Appx. Ind.