Q&A (Frequently Asked Questions)
No. Issue Itemstocheck Possiblecauses→Solutions See
4-5 EOM signal is not
output.
Measured
value
Not updated See No. 3-2 above.
p. 168
EOM signal logic (The EOM signal turns ON once
measurement is completed.)
-
EOM signal
setting
Pulse The pulse output time is short and
the controller cannot detect the
EOM signal.
→ Increase the pulse output time
of the EOM signal or set the output
setting to "hold".
p. 137
Hold The measurement time is short,
and the interval during which the
EOM signal is OFF cannot be
detected.
→ Change the EOM signal output
setting to “pulse”.
4-6 The Hi, IN and Lo
signals are not
output.
Comparator
judgment
results
Are not displayed → See No. 1-3 above.
p. 168