Starting Measurement
Trigger function settings
Trigger source
You can set whether to enable trigger input from an external device. Setting the trigger source to
EXTERNAL enables use of the EXT I/O TRIG terminal as well as the *TRG command. The default
setting is INTERNAL (RUN state).
Delay
You can set the delay time from trigger input until the start of measurement from 0 ms to 9999 ms
in increments of 1 ms. The default setting is PRESET (0 ms).
Adjust the delay time for measurement targets that require some time to respond. Set a long
delay time at rst and then gradually shorten the time while viewing measured values.
Measurement count
You can set how many measurements to perform for each trigger event from 1 to 5000. The default
setting is 1. This setting is disabled while the instrument is in the RUN state.
(Measurement screen) MENU > TRIG > SOURCE
1
1
2
Setting the trigger source
INTERNAL Sets the instrument to the RUN
state (default setting).
EXTERNAL Enables trigger input from an
external device.
2
1
2
3
Settingthedelay
PRESET No delay time (0 ms)
(default setting)
MANUAL Sets the delay time.
˄ Increases the setting by 1.
˅ Decreases the setting by 1.
(Valid setting range: 0 ms to 9999 ms)
3
Measurement