EasyManua.ls Logo

Hioki IR3455 - Page 65

Hioki IR3455
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.1 Pre-Operation Inspection
1
2
3
4
5
6
7
8
9
10
11
付録
索引
61
7. Check to see if the insulation resistance
reading of the instrument is 20 M.
8. Stop insulation resistance measurement.
See 3.2.2 "Ending Measurement" (page 70).
9. Short-circuit the tips of the clips of the
red and black test leads of the instru-
ment.
10. Press the key to see if the test volt-
age setting is [5.00 kV].
11. Hold down the key for more than
one second to start insulation resistance
measurement.
12. Check to see if the insulation resistance
reading of the instrument is 0.00 M.
If a problem exists, discontinue use of
the instrument.
Insulation resistance
Voltage

Table of Contents

Related product manuals