Appendix 16 Calibration
A35
Appendix
When using the YOKOGAWA 2792 to calibration
Use the 4-terminal Lead from Hioki. Note that connection cannot be made with the Clip
Type Lead.
• For more information about 0 calibration connections, see "Appendix 6
Zero Adjustment" (p.A8).
• Adequate noise countermeasures must be implemented during calibration.
Excessive noise will cause measured values to fluctuate or diverge. Con-
nect the metal exterior of standard resistors and dial resistors to the instru-
ment’s GUARD potential.
See: "Appendix 7 Unstable Measured Values" (p.A13)
• Do not use alligator clips with the voltage detection terminals. Thermal
EMFs may cause measured values to diverge.
Correct
4-terminal Lead Clip Type Lead
Wrong