Sensepoint XCD Technical Manual             SPXCDHMANEN Issue 15 
 
48 
 
 
 
 
 
 
Note 1: To comply with IEC EN 60079-29-1 and EN 50104, if the Inhibit current is set to 4 mA, the 
timeout option must be set. The maximum value for the timeout is 15 minutes. 
 
Note 2: To comply with IEC EN 60079-29-1 and EN50104, the alarms must be configured as latching 
by setting the corresponding relay to latching, even if the relay is not being used. If relays are used the 
action must be configured as ‘energized’ for fail safe operation. 
 
Note 3: ‘Init. Baseline’, ‘Select Sensor’ and ‘Select Gas’ are available for flammable/IR sensors only.