EasyManua.ls Logo

Honeywell ST 800 - Page 109

Honeywell ST 800
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Revision 10 ST 800 SmartLine Pressure Transmitters User’s Manual Page 99
MSG
CODE
AGENCY
TYPE OF PROTECTION
COMM
OPTION
Electrical
Parameters
Ambient
Temperature
C
ATEX
Flameproof:
FM12ATEX0029X
II 1/2 G Ex d IIC Ga/Gb
II 2 D Ex tb IIIC T 95
o
C Db
DE/HART
/FF
Note 1
T5: -50 ºC to 85ºC
T6: -50 ºC to 65ºC
Standards: EN 60079-0: 2011 EN 60079-1 : 2007 EN 60079-31 : 2009
EN 60079-26 : 2007 EN 60529 : 2000 + A1
Intrinsically Safe:
Sira12ATEX2233X
II 1 G Ex ia IIC Ga
FISCO Field Device (Only for FF Option)
Ex ia IIC T4
DE/HART
/FF
Note 2
T4: -50
o
C to 70
o
C
Non Sparking:
Sira12ATEX4234X
II 3 G Ex nA IIC Gc
DE/HART
/FF
Note 1
T4: -50
o
C to 85
o
C
Standards: EN 60079-0: 2011 EN 60079-11 : 2011 EN 60079-26 : 2006
EN 60079-15 : 2007 IEC 60529 : 2009 with Corr 3
Enclosure: IP66/ IP67
ALL
ALL
ALL
D
IECEx
Intrinsically Safe:
SIRA 12.0100
Ex ia IIC Ga
FISCO Field Device (Only for FF Option)
Ex ia IIC Ga
DE/HART
/FF
Note 2
T4: -50
o
C to 70
o
C
Non Sparking:
SIRA 12.0100
Ex nA IIC Gc
DE/HART
/FF
Note 1
T4: -50
o
C to 85
o
C
Standards: IEC 60079-0: 2011 IEC 60079-11 : 2011 IEC 60079-26 : 2006
IEC 60079-15 : 2011 IEC 60529 : 2009 with Corr 3
Flameproof:
Ex d IIC Ga/Gb
Ex tb IIIC Db T 95
o
C
DE/HART
/FF
Note 1
T5: -50 ºC to 85ºC
T6: -50 ºC to 65ºC
Standards: IEC 60079-0: 2011 IEC 60079-1 : 2007 IEC 60079-31 : 2008
IEC 60079-26 : 2006 IEC 60529 : 2009 with Corr 3
Enclosure: IP66/ IP67
ALL
ALL
ALL
E
SAEx
(South
Africa)
Intrinsically Safe:
Ex ia IIC Ga
FISCO Field Device (Only for FF Option)
Ex ia IIC Ga
DE/HART
/FF
Note 2
T4: -50
o
C to 70
o
C
Non Sparking:
Ex nA IIC Gc
DE/HART
/FF
Note 1
T4: -50
o
C to 85
o
C
Flameproof:
Ex d IIC Ga/Gb
Ex tb IIIC T 95
o
C Db
DE/HART
/FF
Note 1
T4: -50
o
C to 85
o
C
Enclosure: IP66/ IP67
ALL
ALL
ALL