HP 54501A - Page 217

HP 54501A
386 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMPare
Query
Syntax:
:MEASure:COMPare?
<
measurement
>
Where:
<
measurement
>
:;=
{RISetime
|
FALLtime
|
FREQuency
|
PERiod
|
PWIDth
|
NWIDth
|
VAMPIitude
|
VBASe
|
VTOP
|
VPP
|
VAVerage
|
VMAX
|
VMIN
|
VRMS
|
DUTycycle
|
DELAY}
Returned
Format:
[:MEASure:COMPare] <
measurement
>,
<upperjmit>,
<lower_limit>
<NL>
Example:
DIM
Cmp$[50]
OUTPUT 707;"MEASure:COMPare?
VPP"
ENTER
707;
Cmp$
PRINT
Cmp$
For
example,
the
sequence
required
to
do
a
limit
test
on
frequency
is:
OUTPUT
707;":MEASURE:FREO" !Select
measurement
OUTPUT 707;":MEASURE:COMPARE FREO,1000HZ,10HZ"
'Set
measurement
limits
OUTPUT 707;":MEASURE:LlMITTEST
MEASURE"
"Start
test
Note
The
only
way
to see
if
a
limit
test
failure
has
occurred
over
the
bus
is
by
checking if bit
3
of the
status
byte
is
set
to
a 1.
HP
54501A
Programming
Measure
Subsystem
14-13