POSTfailure
POSTfailure
command/query
The
:MEASURE:POSTFAILURE
command
specifies
what
will
occur
after
a
violation
has
been
found
by
the
limit
test.
If
CONTINUE
is
selected,
the
instrument
will
continue
to
look
for
another
violation.
If
STOP
is
selected
the
instrument
will
stop
the
limit
test.
If
CONTINUE
is
selected and
a
violation
is
found
the
violation
will
be
written to
the
desired
location.
If
a
waveform
memory
is
selected
as
the
destination
then
all
subsequent
violations
will
overwrite
the
previous
violation.
The POSTFAILURE
query
returns
the
current
selection.
Command Syntax:
:MEASure:POSTfailure
{CONTinue
|
STOP}
Example:
OUTPUT 707;":MEAS:
POST"
Query Syntax:
:MEASure:POSTfailure?
Returned
Format:
l:MEASure:POST<ailure] {CONTinue
|
STOP}<NL>
Example:
DIM pf$[50]
OUTPUT 707;":MEASURE:
POSTFAILURE?"
ENTER
707;
Pf$
PRINT
Pf$