RANGe
RANGe
command/query
The
:TIMEBASE:RANGE
command
sets
the
full
scale horizontal
time
in
seconds.
The RANGE
value
is
ten
times
the
time
per
division.
The
RANGE
query
returns
the
current range
value.
Command
Syntax:
:TIMebase:RANGe
<
range
>
<
range
>
::=
20ns
to
50s
in
a
1,2,5
sequence"
Example:
OUTPUT 707;":TIM:RANG 100
MS"
Query Syntax:
:TIMebase:RANGe?
Returned
Format:
[:TIMebase:RANGe] <
range
>
<
NL>
Where:
<
range
>
::
=
20
ns
to 50
s
(exponential
-
NR3
format)
Example:
DIM Rng$[50]
OUTPUT
707;
":TIMEBASE:RANGE?"
ENTER
707;Rng$
PRINT
Rng$