NWIDth
NWIDth
command/query
The
:MEASURE:NWIDTH
command
places the
instrument
in the
continuous
measurement
mode and starts
a
NWIDTH
measurement.
The
NWIDTH
query
measures
and
outputs
the
width
of the
first
negative
pulse
on
screen
using
the
50%
levels with
Standard
measurements
selected.
If
User
Defined
measurements
are
selected,
then
the
measurement
is
made
at
the
mid
threshold
value.
The
algorithm
is:
if
the
first
edge
on
screen
is
rising
then
width
=
(time at second rising
edge
time at
First falling
edge)
else
width
=
(time
at
First
rising
edge
-
time
at
first
falling
edge)
Command
Syntax:
:MEASure:NWiDth
Example:
OUTPUT
707;":MEAS:NW1DTH"
Query
Syntax:
:MEASure:NWiDth?
Returned Format:
[:MEASure:NWiDth] <
value
>
<NL>
Where:
<
value
>
::
=
negative pulse
width
in
seconds
(exponential
-
NR3
format)
Example:
DIM
Nwd$[50]
OUTPUT
707;":MEASURE:NWIDTH?"
ENTER
707;Nwd$
PRINT
Nwd$
HP
54501A
Programming
Measure
Subsystem
14-29