System Measurements
Making System Measurements
5.
Fix-
using an external ampli$er
@weamp):
An external amplifier may be
required for the carrier-to-noise test. See Chapter 8 for more information
about when a preamplifier is required for the carrier-to-noise test. The
carrier-to-noise test calculates the noise contribution of the external
amplifier and reports that correction on the MORE INFO screen of the
carrier-to-noise test.
l Connect the output of the external amplifier to the analyzer input (or
input of coupler).
l Connect the cable TV signal to the input of the external amplifier.
l Press Analyzer Input EXT AMP YES NO to underline YES.
l Press EXT AMP GAIN then enter the gain value of the external
amplifier. The default value is 20
dB
for the HP 85905A preampliEer.
l Press EXT AMP
NZ
FIG then enter the noise figure value of the
external amplifier. The default value is 7
dB
for the HP 85905A
preamplifier.
l Press
Prev
Menu to return to the Setup menu.
The TOTL PWR
Q
INPUT function measures the total RF power into the
analyzer. Use this function to externally adjust the total power into the
analyzer or preamplifier for optimization of measurement dynamic range.