iCR 3600
TM
– Service Manual
7.3.5 Up-side Down Cassette
Symptom: Honeycomb/hexagonal pattern throughout the image.
Cause: Exposing a cassette up-side down.
Solution: First ensure that the cassette is erased, then re-expose the cassette with the
carbon-fiber side facing up, then re-scan.
7.3.6 Grid Lines/Moir
´
e Effect
Symptom: Grid lines/Moir
´
e Effect
Cause: A grid with the wrong LPI, or a misoriented cassette under the grid.
c
2007-2009
Confidential and Proprietary
Property of iCRco, Inc.
32 of 61 Document # 3600-02A Rev C
March 22, 2010