5.5 Schedule of Maintenance
The following is a schedule of maintenance for the iCR 3600.
The following maintenance may be performed by end users:
Maintenance Procedure Frequency
Clean cassettes Weekly
Clean iCR 3600 exterior covers Monthly
The following
i
CR 3600 maintenance must be performed by an
i
CRco authorized service
engineer only:
Maintenance Procedure Frequency
Clean mirrors Yearly
Vacuum inside case Yearly
Clean fan filters Monthly or when visibly dirty.
The following
i
CR 3600 calibration maintenance must be performed by an
i
CRco authorized
service engineer only:
Maintenance Procedure Frequency
Check image performance (Contrast/Noise Ratio and Spacial Resolution) Quarterly
Perform Exposure Index calibration Yearly
5.6 Periodic Cleaning
Periodic cleaning of iCRco products should be done on a monthly basis.
5.6.1 Cleaning the Outside of the iCR 3600
Note
It is important that the covers remain on the
i
CR 3600 at all times. The covers should
only be removed for service by an
i
CRco authorized technician, then immediately replaced.
CAUTION
Do
not
clean the galvo mirror. Dust and fibers in the laser beam path may
affect the radiographic image.
The outside covers of the
i
CR 3600 should be cleaned with a slightly dampened cloth or a
dry cloth moistened with Ball
R
SUNUP
R
glass cleaner or Sprayway
c
glass cleaner.
5.6.2 Cassette Cleaning
CAUTION
At no time should abrasive cleaners or chemicals be used to clean the cassette
or plate.