7.4.3 Clicking Noise Returning to Home
Symptom: Clicking noise heard from CR unit when returning to the home position.
Cause: At the end of the scan: Eraser offset.
At home position: Fuses / sensor.
Solution: At the end of the scan: Change eraser offset.
At home position: Check fuses as cover in Section 8.12. Check the Power Distribution board
/ Move sensor.
7.4.4 Not Finding Home Position
Symptom: The CR unit is not finding the home position.
Cause: Bad fuses, Power Distribution board, or Move sensor.
Solution: Check fuses as cover in Section 8.12. Check the Power Distribution board /
Move sensor.
7.4.5 Motor Gear/Slide Lines
Symptom: Noises while scanning.
Cause: Bad slide / misaligned motor gear
Solution: Listen for noises while making a test scan. Leave the bay door open and watch
the cassette to see if it bumps or shifts during the scan. If this is the case, the unit will either
need to be realigned or have the slide repaired/replaced. Please contact Technical Support,
as motor gear/slide issues are not serviceable in the field.
7.4.6 Cassette Door Stays Open
Symptom: Carbon fiber door will not stay closed.
Cause: Broken catch.