NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
61
9. Option to set/calculate a dilution factor for manual diluted samples.
10. Apply the blank ddH
2
0 or buffer to the illuminated sample window on the
pedestal and close the lid arm for the reference measurement and select
blank to initiate the reading.
Note: The illumination of the sample window can be switched off in the
preferences.
11. Use a lint-free laboratory wipe to clean both the sample window on
pedestal and mirror in lid arm prior to applying the next sample.
Note: It could be helpful to apply the blank a second time and read it as a
sample to ensure a proper blank.
12. Apply sample to the sample window on pedestal and press the sample
button to initiate the measurement.
If auto sample is active the sample measurement is automatically started
after closing the lid arm.