EasyManuals Logo

Implen NanoPhotometer N120 User Manual

Implen NanoPhotometer N120
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
60
Note: A new Blank is recommended when changing the dilution lid.
5. Select the Units in which the concentration should be calculated.
Options are ng/µl (default), µg/µl and µg/ml. pmol/µl if a nucleic acid sequence is entered
for nucleic acid factor calculation.
6. The background correction is enabled at 320 nm by default. Selection options are 320 nm,
340 nm or any wavelength in the range of 220 350 nm. The background correction can
be disabled with the toggle switch.
7. Air bubble recognition is disabled by default. Setting to on detects air
bubbles, lint residues and bad sample conditions of the sample.
Note: Lint residues and bad sample conditions are detected even if the
air bubble recognition is set to off.
8. For dye labelled samples tick the dye labels in the list, which should be used for result
calculation.
Note: If the used dye is not available in the list, please go to preferences and add a custom
dye to the dye list.
There is an option for dye correction which can be enabled/disabled with a toggle switch.
Note: Dye correction is only available for single dye selection.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen NanoPhotometer N120 and is the answer not in the manual?

Implen NanoPhotometer N120 Specifications

General IconGeneral
BrandImplen
ModelNanoPhotometer N120
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals