53
Reset SF
6
Density Trend Data
If an SF
6
density sensor is found to be faulty, the SF
6
Trend
calculation should be re-started after replacing the sensor.
This can be done by resetting the accumulated SF
6
data
for that channel. When this action is taken, a new array
of data will be accumulated and a new SF
6
Trend will be
calculated for that channel. Until there is a full array of data
to calculate the new Trend, the Condence Level for that
channel will indicate less than 100%.
To reset the SF
6
Trend Data for a channel, select the
ACTION tab and click “Reset
SF
6
Trend Data and Alarm
Channel A, B or C”. A yellow bar will appear at the top of
the window.
Click “YES” to conrm the action. The correct
Administrator’s password will need to be given for the
change to be saved. Click “Apply” to complete the action.