Equipment Connections 7-5
kina22e chapter 7.fm Technical Handbook(1206)
Pin assignment (default)
All PLC outputs can be defined by the user from a list of commands. The commands
can be selected in the SETTINGS / INTERFACES / SELECT PLC OUTPUTS
submenu. For details on this submenu please refer to Section 4.6.6.
Leak
If the preset trigger level is exceeded, relay contact NO will close and the relay
contact NC will open respectively.
Ready
As long as the Ecotec E3000 is ready for measurements, the relay contact NO is
closed and the relay contact NC is open respectively..
7.1.4.2 Recorder Outputs
Technical data
Pin assignment
For details please refer to section 4.6.2 (Recorder outputs).
Pin Type of contact Relay Output (default)
10 Currentless closed (NC) Leak
22 Common
23 Currentless open(NO)
24 Currentless closed (NC) Ready
11 Common
12 Currentless open (NO)
Analogue output
0 … 10 V
max. 1mA
Pin Command
1 Leak rate, in linear or logarithmic scale
14 Selected gas number