Loading...
8-1 Maintenance
kina22e chapter 8.fm technical handbook(1206)
8 Maintenance
8.1 Maintenance schedule
Required tools
2 screw drivers, size 2
1 ring spanner wrench, 19 mm (delivered with Ecotec E3000)
8 mm hexagonal screw driver (delivered with Ecotec E3000).
Key for Maintenance Schedule:
I Repair level I Customer
II Repair level II Customer with technical training from INFICON
III Repair level III INFICON service engineer
1 To be replaced every 3 years.
2 Date printed on package is latest installation date.
3 Maximum shelf life: 1 year.
Warning
For all maintenance on the Ecotec E3000, the main power must be disconnected
first.
Required maintenance Sub
Assembly
Material
Description
Part no. Operation hours Repair
level
2000 24.000
Check sinter filter and
replace if necessary
Sniffer tip Sinter Filter for
Sniffertip SL3xx,
SL3000-x (5 pcs.)
200 03 500 X I
Replace if W35 „Flow too
Low“
Sniffer tip Felt for Capillary Filter SL3xx,
SL3000-x (50pcs.)
200 001 116 I
Check internal filter and
replace if necessary
Main Unit Internal Filter (6 pcs.) 200 03 679 X II
Turbo pump
SplitFlow 50
Main Unit Operating Fluid Reservoir 200 003 801 1, 2 II
Bearings
III
Replace diaphragm Diaphragm Pump
MVP015
Wearing parts kit for
diaphragm pump
200 03 504 X III
Clean or replace the air filter
at the bottom of the chassis
Main Unit Air Filter ECOTEC E3000
(104x15 mm, 5 pcs.)
200 001 552 X I
Replace filament (only on
failure)
Main Unit Spare Cathode for
Transpector
200 001 545 III
Replace gas reservoir after 2
years
ECO-Check Replacement gas reservoir 531-010 1, 3

Table of Contents