4-26 Equipment Settings
kina22e chapter 4.fm technical handbook(1206)
4.6.3 RS232 Protocol
The RS232 protocol can be set to “ASCII”, “DIAGNOSTICS”, “PRINTER AUTO” or “PRINTER
MANUAL
”.
The
ASCII PROTOCOL is a protocol similar to the SCPI, a standard protocol widely used
for measuring equipment.
For details on this protocol please refer to the Interface Description (kins22e1).
Diagnostics
The
DIAGNOSTICS protocol is a binary protocol used for analysis in quality control and
trouble shooting during manufacturing and service.
Default setting:
ASCII
Printer Manual
The
PRINTER MANUAL protocol allows to send leak rates on pressing the right sniffer
probe button. One line will be send for each currently enabled gas.
Notice: When in any of the two printer modes, no
ASCII commands must be sent to
the Ecotec E3000 as this will interrupt the printer mode.
When set to I·Guide mode the
PRINTER MANUAL mode is without function.
Printer Auto
In
PRINTER AUTO the leak rate will be sent any time the set trigger level is exceeded.
At the time the signal drops below the trigger level the maximum detected leak rate
will be sent only for the gas(es) that exceeded the trigger.
When set to I·Guide mode the leak rate results will be sent after each test point
together with the test point number.
After the test cycle has been completed, the summarized global leak rate will be sent.
Example:
Format
Date Time Gas name Leak rate Unit
Example
03.05.2005 12:31 R134a 2.34 g/a
03.05.2005 12:31 R600a <0.1 g/a
Format
Date Time Gas name Leak rate Unit
Example
03.05.2005 12:31 R134a 2.34 g/a
Point no. Date Time Leak rate Unit
Point 01 0.1 g/a
Point 02 0.1 g/a