Maintenance 8-14
kina22e chapter 8.fm technical handbook(1206)
The capillary may be removed in that way for the following sniffer tips:
3 Take out the steel capillary and remove the taper gasket at the top of the sniffer tip.
Notice: The capillary can be cleaned with compressed air or a thin steel wire.
4 Reinsert the steel capillary and reinstall the sniffer tip.
5 Screw the plastic capillary filter onto the sniffer tip.
6 Please recalibrate the Ecotec E3000 for most accurate display of leak rate.
8.7.2 Switching from plastic to metal capillary filter
1 Unscrew the plastic capillary filter.
2 Insert taper gasket (Fig. Fig. 8-14/1)
Notice: When you switch from the plastic capillary filter to the metal capillary filter,
do not forget to re-install the taper gasket as otherwise the sniffer line will
be leaky.
3 Screw in metal capillary filter at the end of the sniffer tip.
4 Please recalibrate the Ecotec E3000 for most accurate display of leak rate.
8.8 Replacing the gas reservoir of the
ECO-Check
Notice: Please take off the lid from the new gas reservoir 48 hours before
installation. Due to accumulation of gas in the membrane during storage,
the leak rate will be higher than certified right after opening. Do not use the
new gas reservoir for calibration during this time.
1 Pull out the ECO-Check. The ECO-Check reference leak is fixed with magnetic
holders and can be pulled out easily.