8-13 Maintenance
kina22e chapter 8.fm technical handbook(1206)
8.7.1 Switching from metal to plastic capillary filter
When you switch from the metal capillary filter to the plastic capillary filter you need
to remove the taper gasket. The plastic capillary filter will not fit with the taper gasket
installed.
1 Remove the two Philips screws in the flange of the sniffer tip and remove sniffer tip.
2 Take a small pin or needle (about 0.5 mm) and push the steel capillary out off the tip
from the top.
Fig. 8-14 Pushing out the steel capillary
Fig. 8-15 Capillary sticking out of sniffer tip flange