8-11 Maintenance
kina22e chapter 8.fm technical handbook(1206)
8.6.2 Replacing the felt discs (with water protection tip)
1 Switch off Ecotec E3000!
2 Screw off water protection tip!
3 Push out filter pads and metal grid from the back side.
4 Re-install metal grid at the bottom of water protection tip.
5 Push in two new filter pads (all the way to the bottom of the water protection tip).
6 Switch on Ecotec E3000
Notice: If the flow has changed by more than 30%, a re-calibration of the Ecotec
E3000 will be required and a corresponding warning will be issued by the
software.
7 From the main menu go to HISTORY & MAINTENANCE / CONFIRM MAINTENANCE /
S
NIFFER TIP FILTER and confirm with the ok button on the right side of the display that
the sniffer tip filter has been replaced.
The actual date and time as well as the actual number of operating hours will be
saved and a reminder for maintenance will be issued again after the preset number
of hours. For details see Chapter 4.7.
8 Please recalibrate the Ecotec E3000 for most accurate display of leak rate.
8.6.3 Checking / replacing the sinter filter
Notice: If the flow has changed by more than 30%, a re-calibration of the Ecotec
E3000 will be required and a corresponding warning will be issued by the
software.
1 Switch off the Ecotec E3000.
2 Remove the two Phillips screws and take off the sniffer tip.
Fig. 8-11 Taking off the sniffer tip