Equipment Settings 4-19
kina22e chapter 4.fm technical handbook(1206)
To edit the parameters of a program select the program by scrolling to the
corresponding menu line and open the selected program by pressing the “E
DIT
button.
In the opening
EDIT PROGRAM sub-menu the name of the program and the gas type(s)
selected for this program are stated.
One or two gases from the max of four gases which may be setup in the
EDIT GAS
submenu may be selected.
If only one gas is selected, the other gas will be set to 0.
If two gases are two be used, these must be different gas types (reflected by different
gas numbers).
For each of the two gases a summarized global trigger value corresponding to the
allowable leak rate for the complete unit under test is stored. If only one gas is
selected, the
GLOBAL TRIGGER B line item is left blank.
In addition, the number of locations to be leak tested, the minimum measuring time
for each location as well as the wait time between two leak tests allowed for moving
the sniffer to the next location is stated.
Each parameter can be edited by scrolling to the appropriate line item and selecting
the highlighted item by pressing the
EDIT button.(If only one gas is selected the
GLOBAL TRIGGER B line item will be skipped on scrolling.)
Notice: The I•Guide mode can also be used as a timer signal only. If the number of
points is set to 0, the Ecotec E3000 will prompt for the next location to be
tested continuously without using the global leak rate function.
Each parameter can be edited by scrolling to the appropriate line item and selecting
the highlighted item by pressing the
EDIT button.
Name
Each program can be named with a free text, descriptive name up to 6 characters.
The name can be edited by highlighting the
NAME line and pressing EDIT.
An alphanumerical entry screen will open. Press
OK after entering the desired name
description.
Fig. 4-20 Editing an I•Guide program (two gases selected)