4-22 Equipment Settings
kina22e chapter 4.fm technical handbook(1206)
Leak rate filter
For all normal operation the I•Filter should be selected. I•Filter is an intelligent filtering
algorithm that yields the best results with respect to noise level and stability of leak
rate values specifically developed for use in the Ecotec E3000 leak detector.
Only in cases where the older Ecotec II model is replaced by an Ecotec E3000 and
the Ecotec E3000 is used in a fixed testing system it may be required to stick with the
“
AUTO” or “LOW” filter that was used in the Ecotec II.
Default: I•Filter
Alarm delay
Under conditions with very unstable background conditions it may be favourable to
not issue an alarm immediately but only release an audio alarm if the trigger level has
been exceeded for an extended time.
This
ALARM DELAY can be entered in 1/10 of a second. Values between 0 and 9.9 sec
will be accepted.
Default value: 0.0 sec (i.e. disabled).
Wake up
If you are working in one- or two-shift operation, you may use this function to
automatically wake up the Ecotec E3000 from
SLEEP mode in the morning to allow
for proper warm-up time (please allow for approx. 30 min).
You have to program a time (in hours and minutes, 24 hours) and the Ecotec E3000
will wake up automatically at the pre-programmed time.
A different wake-up time may be programmed for each weekday. If no time is
entered, the Ecotec E3000 will not wake-up automatically at all.
Default value: disabled for all weekdays