4-36 Equipment Settings
kina22e chapter 4.fm technical handbook(1206)
The date and time of calibration
The number of operation hours at time of calibration
The calibration factor
The peak position (as deviation from mass position)
The flow through the sniffer line at the time of calibration
The filament that was in use at the time of calibration (A/B)
The size of the test leak used (external test leak for external calibration and ECO-
check for internal calibration
The argon current at the time of calibration
The current for the gas to be calibrated with test leak and background signal (only
in
ADVANCED mode)
Maintenance interval
In the
MAINTENANCE INTERVAl sub-menu the number of operating hours for the main
Ecotec E3000 in total, the operating hours of the turbo molecular pump and of the
Transpector are listed.
Also all necessary maintenance tasks are displayed and the number of operating
hours / days until they are due for maintenance the next time are stated.
When the number of operating hours have expired a warning will be issued as a
reminder to perform the appropriate task.
The reminders are issued for the TMP, the fore pump (diaphragm pump) and the air
filter of the main unit.
Please also refer to Section 4.1 (Error Messages and Warnings)!
Fig. 4-27 Example of a
MAINTENANCE INTERVAL menu page