4-38 Equipment Settings
kina22e chapter 4.fm technical handbook(1206)
Replace ECO-Check
(in ADVANCED user mode only)
When replacing the gas reservoir of the ECO-Check reference leak, the new serial
number of the replacement gas reservoir as well as a code containing the calibration
data of the new reservoir needs to be entered in this submenu.
For details on how to replace the gas reservoir of the ECO-Check reference leak
please refer to Section 8.8.
Notice: This menu item will only be available as long as the ECO-Check is enabled
in the
SETTINGS / INTERFACE / ECO-CHECK submenu.
Sniffer tip filter
(in STANDARD user mode only)
The maintenance interval for the sniffer tip filter depends on the operating
environment and may be set to values between 10 and 999 hours.
If set to 0 the reminder message for the sniffer tip filter will be disabled.
Default value: 100 hours
For details on how to perform the maintenance tasks see Chapter 8, Maintenance.
Select Filament
In this submenu you can manually select which Transpector
TM
filament to use.
Default: A
Fig. 4-30 Setting the maintenance interval for the sniffer tip filter and confirming its
replacement