7 Warnings and error messages 57
Translation of the original operating instructions Ecotec E3000, kina22en1-19, 1507
W35 Flow through capillary is too low
(The lower limit value is fallen short of
longer than 2 s.)
The lower limit value is set too high. Lower the lower flow rate limit value, see page 32.
The capillary filter in the sniffer line is clogged. Replace the capillary filter, see page 72.
Confirm the work, see page 65.
Recalibrate the device, see page 37.
The sinter filter in the sniffer handle is clogged. Replace the sinter filter, see page 73.
Confirm the work, see page 65.
Recalibrate the device, see page 37.
The capillary is clogged. Replace the sniffer line.
Recalibrate the device, see page 37.
Replace the multi-function cable of the sniffer line. *
Recalibrate the device, see page 37.
The internal filters in the main unit are clogged Replace the internal filters (three pieces).*
Confirm the work in the service menu.
Recalibrate the device, see page 37.
E36 Flow through capillary too high.
(The upper flow rate limit value is
exceeded longer than 2 s.)
The sniffer line is incorrectly connected. Check the connection.
The upper limit value is set too low. Set the upper limit value for the flow higher, see
page 32.
The capillary is broken or leaking. Replace the sniffer line.
Recalibrate the device, see page 37.
Replace the multi-function cable of the sniffer
line.Replace the multi-function cable of the sniffer
line.*
Recalibrate the device, see page 37.
E37 Capillary broken! (>10 s)
(The upper limit value is exceeded longer
than 10 s. The emission is switched off to
protect the cathodes.)
The upper limit value is set too low. Set the upper limit value for the flow higher, see
page 32.
The capillary is broken or leaking. Replace the sniffer line.
Recalibrate the device, see page 37.
Replace the multi-function cable of the sniffer
line.Replace the multi-function cable of the sniffer
line.*
Recalibrate the device, see page 37.
E38 Capillary broken! (>60 s)
(The upper limit value is exceeded longer
than 60 s. The pumps are switched off to
protect the cathodes.)
The upper limit value is set too low. Set the upper limit value for the flow higher, see
page 32.
The capillary is broken or leaking. Replace the sniffer line.
Recalibrate the device, see page 37.
Replace the multi-function cable of the sniffer
line.Replace the multi-function cable of the sniffer
line.*
Recalibrate the device, see page 37.
E39 Emission failed
(The emission on both cathodes cannot
be switched on.)
If the device was switched off for a longer period
of time, this error can occur during the first
10 minutes after start-up.
Both cathodes are defective.
The Transpector is defective.
Confirm the error message and restart the device. If the
problem still exists: Contact the INFICON customer
service.
Table 10: Warning and malfunction messages (cont.)
No. Message Possible error sources Fault clearance

Table of Contents

Other manuals for Inficon Ecotec E3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon Ecotec E3000 and is the answer not in the manual?

Inficon Ecotec E3000 Specifications

General IconGeneral
BrandInficon
ModelEcotec E3000
CategorySecurity Sensors
LanguageEnglish

Related product manuals