Source Mode Operation 
  Copyright ©ITECH Electronic Co., Ltd.  43 
5.3.3 AC+DC Mode 
If the output mode select to AC+DC Mode, the instrument will simulate AC and 
DC power supply, which can add DC component to AC voltage. Set the output 
voltage in the main interface, as shown in the figure below. 
 
 
Under AC+DC  mode,  Set  the  output voltage  and  frequency of  the AC  power 
supply in the main interface, and set the DC voltage in the configure menu. 
Vac: you can set the Vac under the main interface or under the config interface. 
DC: set the DC component under the config interface. The setting range is 10% 
of rated voltage. 
  NOTE 
The AC + DC mode not only expands the application range of pure AC voltage, but also expands 
the application range of DC component in laboratory testing.When using AC+DC for testing, please 
first understand the ripple parameters of the instrument when it is used as a DC power supply. If 
there are strict noise requirements, additional DC noise filters are needed to obtain low noise and 
good stable DC voltage for testing. 
 
5.3.4 DC+AC Mode 
If the output mode select to DC+AC Mode, the instrument will simulate DC and 
AC power supply, which can add AC component to DC voltage. Set the output 
voltage in the main interface, as shown in the figure below.