Source Mode Operation 
  Copyright ©ITECH Electronic Co., Ltd.  54 
 
 
List parameters description: 
Start phase setting, displayed in AC mode. 
By  default,  the  program  automatically  calculates 
this  value  (which  is  guaranteed  to  be  continuous 
with  the  previous  step).  If  the  user  modifies  this 
parameter,  the  user  set  value  is  used  without 
guarantee of continuity. 
Phase difference between ABC, only displays in AC 
3-phase mode. 
Frequency and slope, displayed in AC mode. 
Waveform  type,  every  basic  waveform  can  be 
selected, displayed in AC mode. 
Method of step jumps to next step.   
Time: when the time is out, jumps to next step 
Trig: receive a trigger signal, jumps to next step. 
Phase: jumps to next step at this phase. 
Whether outputs a signal when this step is ending. 
 
4.  Press [Esc] to return Edit interface, Press [Save]. 
At list Edit interface, click the step number, the [Insert]/[Paste]/[Cut]/[Copy] 
/[Delete] will display, click the key to edit. 
5.  Press [Esc] to return. 
5.12.2 Select/Run List File   
If several List files are edited, press  Recall to recall the List file to be tested. 
Detailed operation steps are as below: 
1.  Press  [Shift]+[V-set](list)  on  the  front  panel  to  enter  the  List  function 
configuration interface. 
2.  Press [Open], select the saved List01 csv file, and press [Enter] to enter 
the file.   
3.  Press [On/Off] on the front panel, turn on the output. 
4.  Press [Run] in the list function interface.