SCPI subsystem commands
Detailed information an all SCPI commands are covered in the User’s Manual. In this section,
only the command used for basic operation are summarized.
Defaults
:SYSTem
SYSTem Subsystem
:PREset
Set the 2002 to the system present defaults.
Some of the selected defaults include: DCV, continu-
ous triggers, normal speed (1.0 PLC), auto range and
auto filter.
:POSetup <name>
Select power-on setups:
RST
, PRESet or SAV0 through
SAV9. Instrument will power up to the *RST or system
present defaults, or one of the
*SAV0 – *SAV9 setups.
Fresh readings
[:SENSe[1]] SENSe Subsystem
:DATA:FRESh? Requests a “fresh” reading. This command will not re-
quest the same reading twice. If a new reading has al-
ready been triggered, this command will wait until the
new reading is available.
:CALCulate[1] CALCulate Subsystem
:DATA:FRESh? Requests a “fresh” calculated reading. When CALC1
is enabled, sens:data:fres? is the reading before calcu-
lation and calc:data:fres? is the reading after calcula-
tion. When CALC1 is disabled, the readings are the
same.
Function configuration
[:SENSe[1]] SENSe Subsystem
:FUNCtion <name> Select measurement function:
<name> = 'VOLTage:DC' DCV function
'VOLTage:AC' ACV function
'CURRent:DC' DCI function
'CURRent:AC' ACI function
'RESistance' Ω2 function
'FRESistance' Ω4 function
'FREQuency' FREQ function
'TEMPerature' TEMP function