EasyManuals Logo

Keithley 3706A Reference Manual

Keithley 3706A
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #530 background imageLoading...
Page #530 background image
Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-151
An error is generated if:
• You try to set dmm.autozero for "continuity" or "nofunction". Error code 1114, "Setting
conflicts with function selected," is generated.
• You query dmm.autozero for "continuity" or "nofunction". nil is returned with error
code 1114, "Setting conflicts with function selected."
Example
dmm.func = "dcvolts"
dmm.autozero = dmm.ON
dmm.autozero = dmm.AUTOZERO_ONCE
print(dmm.autozero)
timer.reset()
dmm.autozero=2
time=timer.measure.t()
print(time)
Enables autozero for DC volts.
Refreshes the reference points once and sets
autozero to dmm.OFF or 0.
Output:
0.000000000e+00
Determines the time delay when autozero is selected.
Also see
Autozero (on page 4-3)
dmm.configure.recall() (on page 11-165)
dmm.configure.set() (on page 11-166)
dmm.func (on page 11-179)
dmm.nplc (on page 11-209)
dmm.reset() (on page 11-219)
reset() (on page 11-308)

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A

Related product manuals