EasyManuals Logo

Keithley 6487 Reference Manual

Keithley 6487
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #186 background imageLoading...
Page #186 background image
11-2 Common Commands Model 6487 Reference Manual
Common commands
Common commands (summarized in Table 11-1) are device commands that are common
to all devices on the bus. These commands are designated and defined by the IEEE-488.2
standard.
Table 11-1
IEEE-488.2 common commands and queries
Mnemonic Name Description Ref
*CLS Clear status Clears all event registers and error queue. Section 10
*ESE <NRf> Event enable command Program the standard event enable register. Section 10
*ESE? Event enable query Read the standard event enable register. Section 10
*ESR? Event status register query Read the standard event enable register and clear it. Section 10
*IDN? Identification query Returns the manufacturer, model number, serial
number, and firmware revision levels of the unit.
A
*OPC Operation complete command Set the operation complete bit in the standard event
register after all pending commands have been
executed.
B
*OPC? Operation complete query Places an ASCII “1” into the output queue when all
pending selected device operations have been
completed.
B
*OPT? Option query Returns model number of any installed options.
*RCL <NRf> Recall command Returns Model 6487 to the user-saved setup. C
*RST Reset command Returns Model 6487 to the *RST default
conditions.
D
*SAV <NRf> Save command Saves the present setup as the user-saved setup. C
*SRE <NRf> Service request enable command Programs the service request enable register. Section 10
*SRE? Service request enable query Reads the service request enable register. Section 10
*STB? Status byte query Reads the status byte register. Section 10
*TRG Trigger command Sends a bus trigger to Model 6487. E
*TST? Self-test query Performs a checksum test on ROM and returns the
result.
F
*WAI Wait-to-continue command Wait until all previous commands are executed. G

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%