9 Verification
General introduction:
According to EU directive 90/384/EEC or 2009/23EG balances must be officially verified
if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purposes
d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard.
Verification notes:
An EU type approval exists for balances described in their technical data as verifiable.
If a balance is used where obligation to verify exists as described above, it must be
verified and re-verified at regular intervals.
Re-verification of a balance is carried out according to the respective national
regulations. The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
Verification of the balance is invalid without the seal.
The seal marks attached on verified balances point out that the balance may
only be opened and serviced by trained and authorised specialist staff. If the
seal mark is destroyed, verification looses its validity. Please observe all
national laws and legal regulations. In Germany a re-verification will be
necessary.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ABS-N Series and is the answer not in the manual?

KERN ABS-N Series Specifications

General IconGeneral
BrandKERN
ModelABS-N Series
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals