ABS-N_ABJ-NM-BA-e-1715 83
16.3.1 Selection of a standard setting „MODE 1 - MODE 5“
1. Invoke Data Output menu
Press PRINT for 3 sec. in weighing mode.
2. Select function
Press the navigation buttons ( ) repeatedly
until „COMM.SET“ is displayed.
Confirm with PRINT
Press the navigation buttons ( ) repeatedly until the
desired mode „MODE 1 - MODE 5“ is displayed. The
current setting is marked by the stability display ( ).
with ( ) on
without ( ) off
Change using TARE.
3. Return to weighing mode
Press ON/OFF repeatedly or for 3 s.