96 ABS-N_ABJ-NM-BA-e-1715
18.1 Error messages
Error message
Explication
Remedy
ERR H Hardware error
Switch balance off and on
again. If the error message
remains inform manufacturer.
ERR C
High zero point shift during
adjustment
Use ON/OFF to return into
weighing mode. Restart
adjustment process.
Objects present on weighing
plate.
Missing weighing plate
CAL D Display instable
Check ambient conditions
(draught, vibrations etc.) Use
ON/OFF to return into weighing
mode. Restart adjustment
process.
ERR N
Error at numerical input (e. g.
wrong password)
Correct input
ERR W Wrong application
The balance changes to the
previous state. Correct
application.
COM ERR Wrong remote control order.
The balance changes to the
previous state. Correct remote
control order.
- OL Missing weighing plate Install weighing plate correctly
OL Overload Reduce load
ABORT Process cancelled
WAIT Wait for process
BUSY
When adjustment was started
objects were present on
weighing plate
Remove object and continue
adjustment
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ABS-N Series and is the answer not in the manual?

KERN ABS-N Series Specifications

General IconGeneral
BrandKERN
ModelABS-N Series
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals