50 ABS-N_ABJ-NM-BA-e-1715
12.2 Auto zero function
The auto zero function is not available when formula mode is enabled
(see chapter 14.3).
1. Call up menu
Press TARE for 3 sec. in weighing mode.
2. Select function
Press the navigation buttons ( ) repeatedly until
A.ZERO“ is displayed. The current setting is marked by
the stability display ( ).
Further steps depend on the desired setting:
Stability
display ( )
Function set/update cancel
yes
ON
Press PRINT and
continue with step 3.
Press TARE and
continue with step 4.
no
off
Press TARE and
continue with step 3.
continue with step 4.
3. Define zero range
Press TARE
Use the navigation keys ( ) to enter the zero range
see chpt. 3.1.1 „Numeric input“.
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
When entering the zero range observe the currently set weighing unit.
If the weighing unit is changed at a later moment, the zero range must be
adapted to the new weighing unit (step 3.).
Upper limit zero range: 99 d (in the displayed weighing unit)
Lower limit zero range: 1d (in the displayed weighing unit)
Example for a balance d = 0.0001 g
Unit
Lower limit
Upper limit
g
0.0001 g
0.0099 g
ct
0.001 ct
0.099 ct

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ABS-N Series and is the answer not in the manual?

KERN ABS-N Series Specifications

General IconGeneral
BrandKERN
ModelABS-N Series
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals