EasyManua.ls Logo

KERN ABS-N Series - Page 165

KERN ABS-N Series
287 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ABS-N_ABJ-NM-BA-e-1715 69
0. Call up menu
In weighing mode press the MENU button twice
1. Select function
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT
Press the navigation buttons ( ) repeatedly
until „CHECK.W“ is displayed. The current setting is
marked by the stability display ( ).
Further steps depend on the desired setting:
Stability
display ( )
Function set/update cancel
yes
ON
Press PRINT and
continue with step 3.
Press TARE and
continue with step 4.
no
off
Press TARE and
continue with step 3.
continue with step 4.
2. Set limit values
When entering the limit values ensure that the values match logically one
with another, i.e. the lower limit value must not be greater than the upper
one.
If not considered, the balance will adapt the limit values automatically.
Press the navigation buttons ( ) repeatedly until
HI.LIM is displayed.
Press TARE. The current setting will be displayed.
Use the navigation buttons ( ) to enter the desired
value, see chpt. 3.1.1 „Numeric input“.

Table of Contents

Related product manuals