Page #165 background imageLoading...
Page #165 background image
ABS-N_ABJ-NM-BA-e-1715 69
0. Call up menu
In weighing mode press the MENU button twice
1. Select function
Press the navigation buttons ( ) repeatedly until
TOOLS“ is displayed.
Confirm with PRINT
Press the navigation buttons ( ) repeatedly
until „CHECK.W“ is displayed. The current setting is
marked by the stability display ( ).
Further steps depend on the desired setting:
Stability
display ( )
Function set/update cancel
yes
ON
Press PRINT and
continue with step 3.
Press TARE and
continue with step 4.
no
off
Press TARE and
continue with step 3.
continue with step 4.
2. Set limit values
When entering the limit values ensure that the values match logically one
with another, i.e. the lower limit value must not be greater than the upper
one.
If not considered, the balance will adapt the limit values automatically.
Press the navigation buttons ( ) repeatedly until
HI.LIM is displayed.
Press TARE. The current setting will be displayed.
Use the navigation buttons ( ) to enter the desired
value, see chpt. 3.1.1 „Numeric input“.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ABS-N Series and is the answer not in the manual?

KERN ABS-N Series Specifications

General IconGeneral
BrandKERN
ModelABS-N Series
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals