3. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
4. Start tolerance control
If necessary, place an empty container on the balance and tare it by pressing TARE.
Place weighed goods and wait until the indicator
,
or
appears. With
the help of the indicator check if the weighed goods are under, inside or over the
default tolerance
Example of entry: HI.LIM 7.0000 g
LO.LIM 6.0000 g
UND.RNG 5.0000 g
OVR.RNG 8.0000 g
Sample weight < UND.RG
(sample weight < 5 .0000g)
No indicator is
displayed
UND.RG ≤ sample weight < LO.LIM
(sample weight 5.0000g - 5.9999g)
is displayed
LO.LIM ≤ sample weight ≤ HI.LM
(sample weight 6.0000g - 7.000g)
is displayed
HI.LIM < sample weight ≤ OVR.RNG
(sample weight 7.0001g - 8.0000g)
is displayed
Sample weight > OVR.RNG
(sample weight > 8.0000 g)
No indicator is
displayed