EasyManua.ls Logo

Keysight Technologies 33210A - Page 99

Keysight Technologies 33210A
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Procedures 4
Keysight 33210A Service Guide 99
3 Using the numeric keypad, adjust the displayed amplitude at each setup to
match the measured amplitude (in dBm). Then select ENTER VALUE.
4 After performing setup 62:
a If your calibration procedures require you to verify the adjustment just
made, exit the calibration menu and perform “+10 dB Range Flatness
Verificationon page 74.
b If you are making all the adjustments and then verifying the instrument’s
performance, continue with the next procedure in this chapter.
56
[a]
500 kHz 0.9 Vrms 12 dBm Flatness for 10 dB, Elliptical Filter
57
[a]
1.5 MHz 0.9 Vrms 12 dBm Flatness for 10 dB, Elliptical Filter
58
[a]
3.0 MHz 0.9 Vrms 12 dBm Flatness for 10 dB, Elliptical Filter
59 4.0 MHz 0.9 Vrms 12 dBm Flatness for 10 dB, Elliptical Filter
60
[a]
6.0 MHz 0.9 Vrms 12 dBm Flatness for 10 dB, Elliptical Filter
61
[a]
8.0 MHz 0.9 Vrms 12 dBm Flatness for 10 dB, Elliptical Filter
62
[a]
9.9 MHz 0.9 Vrms 12 dBm Flatness for 10 dB, Elliptical Filter
[a] Constants are stored after completing this setup.
Setup
Nominal Signal
Frequency Amplitude
NOTE
In order to get dBm you must use the numeric keypad (not the knob) to enter the
number, and then select “dBm”.

Table of Contents

Other manuals for Keysight Technologies 33210A

Related product manuals