Keysight E5260/E5270 Programming Guide, Edition 4 4- 11
Command Reference
Command Parameters
Table 4-3 Current Measurement Ranging Type
range
a
a. If the measurement channel forces current, the channel uses the current output range regardless
of the range value.
Ranging type
for measurement mode
without pulse
for measurement mode
that uses pulse
0 Auto ranging Measurement channel uses
the minimum range that
covers the compliance
value.
8, for E5287A+E5288A ASU 1 pA limited auto ranging
9, for E5287A 10 pA limited auto ranging
10, for E5287A 100 pA limited auto ranging
11, for E5280B/E5281B/E5287A 1 nA limited auto ranging
12, for E5280B/E5281B/E5287A 10 nA limited auto ranging
13 100 nA limited auto ranging
14 1 μA limited auto ranging
15 10 μA limited auto ranging
16 100 μA limited auto ranging
17 1 mA limited auto ranging
18 10 mA limited auto ranging
19 100 mA limited auto ranging
20, for E5291A 200 mA limited auto ranging
20, for E5280B/E5290A 1 A limited auto ranging
-8, for E5287A+E5288A ASU 1 pA range fixed
-9, for E5287A 10 pA range fixed
-10, for E5287A 100 pA range fixed
-11, for E5280B/E5281B/E5287A 1 nA range fixed
-12, for E5280B/E5281B/E5287A 10 nA range fixed
-13 100 nA range fixed
-14 1 μA range fixed
-15 10 μA range fixed
-16 100 μA range fixed
-17 1 mA range fixed
-18 10 mA range fixed
-19 100 mA range fixed
-20, for E5291A 200 mA range fixed
-20, for E5280B/E5290A 1 A range fixed